Abstract

We report a case of stent fracture and collapse of balloon-expandable stents caused by shiatsu massage. A 76-year-old man presented with complaints of intermittent claudication of the right lower extremity. Stenoses of the bilateral common iliac arteries (CIAs) were detected. Balloon-expandable stents were deployed in both CIAs, resulting in resolution of symptoms. Five months later, pelvis x-ray showed collapse of both stents. Despite the stent collapse, the patient was asymptomatic, and his ankle brachial index values were within the normal range. Further history showed that the patient underwent daily shiatsu therapy in the umbilical region, which may have triggered collapse of the stent. Physicians should advise patients to avoid compression of the abdominal wall after implantation of a stent in the iliac artery.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.