Abstract

BackgroundTo assess the accuracy of fractionated stereotactic radiotherapy (FSRT) using a stereotactic mask fixation system.Patients and MethodsSixteen patients treated with FSRT were involved in the study. A commercial stereotactic mask fixation system (BrainLAB AG) was used for patient immobilization. Serial CT scans obtained before and during FSRT were used to assess the accuracy of patient immobilization by comparing the isocenter position. Daily portal imaging were acquired to establish day to day patient position variation. Displacement errors along the different directions were calculated as combination of systematic and random errors.ResultsThe mean isocenter displacements based on localization and verification CT imaging were 0.1 mm (SD 0.3 mm) in the lateral direction, 0.1 mm (SD 0.4 mm) in the anteroposterior, and 0.3 mm (SD 0.4 mm) in craniocaudal direction. The mean 3D displacement was 0.5 mm (SD 0.4 mm), being maximum 1.4 mm. No significant differences were found during the treatment (P = 0.4). The overall isocenter displacement as calculated by 456 anterior and lateral portal images were 0.3 mm (SD 0.9 mm) in the mediolateral direction, -0.2 mm (SD 1 mm) in the anteroposterior direction, and 0.2 mm (SD 1.1 mm) in the craniocaudal direction. The largest displacement of 2.7 mm was seen in the cranio-caudal direction, with 95% of displacements < 2 mm in any direction.ConclusionsThe results indicate that the setup error of the presented mask system evaluated by CT verification scans and portal imaging are minimal. Reproducibility of the isocenter position is in the best range of positioning reproducibility reported for other stereotactic systems.

Highlights

  • Stereotactic radiation techniques in form of radiosurgery (SRS) or fractionated stereotactic radiotherapy (FSRT) are frequently employed in patients with skull base tumors in order to increase the precision of radiotherapy and decrease the potential long-term toxicity of treatment [1,2,3].FSRT using a commercially available stereotactic mask fixation system (BrainLAB AG) has been routinely used at University Hospital Sant’Andrea in patients with skull base tumors since 2006

  • The results indicate that the setup error of the presented mask system evaluated by CT verification scans and portal imaging are minimal

  • CT verification Sixteen patients were evaluated in the study for a total of 64 verification CT scans

Read more

Summary

Introduction

Stereotactic radiation techniques in form of radiosurgery (SRS) or fractionated stereotactic radiotherapy (FSRT) are frequently employed in patients with skull base tumors in order to increase the precision of radiotherapy and decrease the potential long-term toxicity of treatment [1,2,3]. FSRT using a commercially available stereotactic mask fixation system (BrainLAB AG) has been routinely used at University Hospital Sant’Andrea in patients with skull base tumors since 2006. An essential prerequisite of a frameless system is that patient fixation and positioning are performed with a high degree of accuracy in order to delivery a safe therapeutic radiation dose. Accuracy of patient positioning reproducibility with a stereotactic mask fixation system using both CT and portal images as in current use in our department is reported and discussed. To assess the accuracy of fractionated stereotactic radiotherapy (FSRT) using a stereotactic mask fixation system

Methods
Results
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.