Abstract

Fractal analysis of digitized images has been investigated in recent years as a potential measure of structural bone strength. Several technical issues associated with such measurements are assessed. In a series of experiments using a hand phantom, the effects of system noise and modulation transfer function on fractal dimension were explored. The authors evaluated a method for correcting the estimated power spectrum using a step-wedge image exposed and digitized under identical conditions as a reference. System noise and modulation transfer function significantly affect estimated fractal dimension in bony regions computed from conventional radiographs. Before conventional radiographs are used for fractal analysis in the clinical environment, many of the technical problems associated with this methodology must be addressed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.