Abstract

A typical echograph transmits ultrasound signals and receives backscattered echoes through several tens of small transducers that compose the array of the probe. The system should process the acquired echoes in a very short time, typically few tens of ms, so that the medical doctor does not perceive any delay between the probe movements and the image update. This implies, especially when high-frame rate is required, the real-time processing of several GB of data per second: a huge constraint that is even emphasized in the new advanced ultrasound modalities investigated nowadays. In this paper, we present a Field Programmable Gate Array architecture which uses a combination of parallel and serial strategies capable to achieve a throughput rate of 3.5 GB/s in ultrasound data processing. The proposed architecture is implemented in the research ultrasound system ULA-OP 256. The performance of this approach is demonstrated by the real-time implementation of Plane Wave Vector Doppler, a novel investigation method that shows the blood velocity fields over a 2D section of an artery. Tests in a carotid artery of a volunteer are presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.