Abstract

To report the results of a novel surgical four-point transscleral suture fixation of intraocular lens (IOL) with four hollow haptics using the double-suture technique. We retrospectively reviewed the medical records of 15 eyes of 15 patients who underwent 4-point transscleral suture fixation of a foldable IOL using the double-suture technique. Preoperative data and follow-up data for at least 4 months were collected for all patients. The IOLs were fixed and centered well. The mean preoperative corrected distance visual acuity was 0.70 ± 0.54 logarithm of the minimum angle of resolution (Snellen 20/102), and it improved to 0.29 ± 0.26 logarithm of the minimum angle of resolution (Snellen 20/39) at the final follow-up ( P = 0.001). No vitreous hemorrhage, hypotony, suture breakage, retinal detachment, IOL dislocation, and iris capture was detected during the follow-up period in any of the patients. We have developed a novel technique for 4-point transscleral suture fixation of IOL using the double-suture technique with 9-0 polypropylene suture. This technique seemed to be safe and it may not require the surgeon to learn any new technique.