On the basis of general theories of Fourier self-deconvolution and curve fitting, an objective procedure to obtain the deconvolution parameters halfwidth, σ, and breaking point, L, is described. In the curve fitting routine, the analytical accessible line shape functions, derived for different apodization functions, are used to simulate the deconvoluted spectra. The fully automated deconvolution program was applied to the IR spectra of the amide I band of dry protein films to determine the secondary structure content.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.