Abstract

We establish an encoding figure of merit for a detector-noise limited Fourier transform spectrometer (FTS) and compare it to the comparable figure for a Hadamard transform spectrometer (HTS). If N measurements are made to establish N spectral densities, the mean square errors obtained with the Fourier system are a factor of 2 greater than for the analogous Hadamard system. The limitation of the Fourier system is partly that it does not truly Fourier analyze the radiation. Instead a cosine squared modulation is imposed on the different spectral frequencies. An additional difficulty is that neither the cosine nor the cosine squared functions form an orthonormal set. This makes the Fellgett's advantage (root-mean-squared figure of merit) for a single detector Michelson interferometer a factor of (N/8)(1/2) greater than for a conventional grating instrument-rather than (N/2)(1/2) as maintained in standard texts. The theoretical limit, which may not be realizable with practical instruments, would be (N)(1/2).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.