Abstract

BackgroundTo evaluate the feasibility of creating flanges using an optic piercing technique with a 6 − 0 polypropylene monofilament for scleral fixation of dislocated one-piece diffractive multifocal intraocular lenses (IOLs).Study DesignExperimental study and case series.SubjectsOptical bench test and eyes with IOL dislocation.MethodsTwo separate 6 − 0 polypropylenes were penetrated twice at the opposite peripheral optic of the TECNIS Synergy IOL (Johnson & Johnson Vision). The root mean square of the modulation transfer function (MTFRMS), at between + 1.00 and − 4.00 D of defocus, was measured in the TECNIS Synergy IOL both with and without optic piercing in the optical bench study. This case series included three eyes from two patients who underwent scleral-fixation of multifocal IOLs using the four-flanged polypropylene optic piercing technique. The postoperative corrected distance visual acuity (CDVA) at 4 m, the uncorrected near visual acuity (UNVA) at 40 cm, and IOL centration were evaluated.ResultsThe optical bench test showed no differences in MTFRMS values measured in the TECNIS Synergy IOL, either with or without optic piercing at all defocuses. In all three case series, the postoperative CDVA at 4 m was 20/20 and UNVA at 40 cm was J1. Postoperative anterior segment photographs showed good centration of IOLs in all cases.ConclusionThe four-flanged polypropylene optic piercing technique for multifocal IOL scleral fixation can provide excellent clinical outcomes and IOL stability after surgery without diminishing the performance of the multifocal IOLs.