Abstract

A four-core optical fiber is demonstrated as a calorimetric gauge for investigation of one-dimensional heat transfer measurements. Transient heat pulses from a Nd:YAG laser of 600 ms duration with a repetition rate of the order of 10 s are delivered onto the cleaved distal end face of the four-core fiber, aiming at one of the single cores only, which cause an optical path length difference between four guiding cores due to the temperature-induced change in the index of refraction and physical length of the targeted fiber core of concern. This results in a shift in the fringe pattern, which is operated in the reflection scheme. A phase shift of 0.43±0.015 rad is measured with a CMOS camera for 40 mW pulses. The thermal heat diffusion length in the selected fiber core is determined to be 2.8 mm, which contains 10.9±0.38 kJ/m2s heat, causing a temperature rise of 1.43±0.05 K.