Abstract

We have studied the effect of plasma treatment on the rectification property and performance of the CdTe radiation detector with the Al Schottky electrode. The Te-rich layer on the CdTe surface etched with Br-methanol caused the degradation of the rectification property of the Al/CdTe Schottky contact. To remove the Te-rich layer, plasma treatment was carried out. The plasma treatment did not roughen the CdTe surface, and it removed the Te-rich layer. In terms of current–voltage characteristics of the Al/CdTe Schottky contact, the leakage current of the samples with plasma treatment was lower than that of the samples without plasma treatment. Moreover, in terms of detector performance, the samples with plasma treatment showed a higher energy resolution than those without plasma treatment. We achieved a high energy resolution of 1.6 keV FWHM at 59.5 keV using the plasma-treated Al/CdTe/Pt detector, which is comparable to the value obtained using a conventional Schottky-type In/CdTe/Pt detector.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.