Abstract

This paper discusses the application of formal methods software engineering (FMSE) to the development of the Computer Automated Resuscitation A (CARA) medical device at Walter Reed Army Institute of Research. Because this system is potentially life critical, a high level of quality was required. A formal engineering approach to the software development activities was chosen to satisfy this need. Specifically, a technique called sequence enumeration was applied to elicit and refine requirements while deriving a formal specification. The fundamentals of the specification process that was used on the project are described along with a brief summary of the project experience in the development and testing phases. The project employed recent advances in Cleanroom software engineering methods along with older box-structured development and usage-model-based statistical testing techniques.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.