Abstract
Following a critical incident on our labour ward, we investigated the forces required to remove different bougies from different tracheal tubes and the effect of lubrication on ease of removal. Two studies were conducted: firstly examining the differences between six different bougies with a standard tracheal tube, and secondly examining the differences between four different tracheal tubes with a standard bougie. The forces varied amongst both the different bougies (p < 0.0001) and the different tracheal tubes (p < 0.0001). Removal was generally easier with lubrication but when corrected for multiple comparisons this did not reach statistical significance.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.