Abstract

This study represent the influence of relative zero error on force proving instruments classification. It is directed to compare the relative zero error resulted from the three loading scenarios applied during calibration of force proving instruments in increasing and decreasing scheme as per ISO 376:2011. It is conclude that the maximum relative zero error resulted after the first loading series at zero position. Two methods are recommended to be proposed as a unique method to calculate the maximum relative error of zero from the first and second loading series or from the preloading series

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.