Abstract

Bone conduction measurements are realized for determination of the threshold value of the cranial bone. The Artificial Mastoid as described in IEC 60373 is a mechanical coupler used for establishment of traceability in bone conduction measurements. Artificial Mastoid must be characterized depending on its force sensitivity. This will guarantee the traceability to national and international standards. A semi-automatic system which was constructed at at TUBITAK UME for characterization of artificial mastoid and uncertainty in measurements will be presented in this paper. The uncertainty in force measurements is stated as ± 0.25 dB between 100 Hz and 8 kHz.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.