Abstract
Abstract Electroencephalography (EEG) and transcranial electric stimulation (TES) require caps for holding the respective electrodes in place. To support the optimal design of such caps, knowledge of the force-displacement curves for each electrode position is desirable. We propose a calibrated setup to traceably measure force-displacement curves which consists of a human head model, a force sensor, a linear guide, a stepper motor, and a multiplexing multimeter. Repeated measures of a textile EEG-cap and a TES-cap show significant non-linearity and hysteresis effects for the force-displacement curves. Our setup will allow for the assessment of the fit of EEG and TES-caps for various head shapes and sizes.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.