Abstract

We have studied the dynamics of quartz tuning fork resonators used in atomic force microscopy taking into account the mechanical energy dissipation through the attachment of the tuning fork base. We find that the tuning fork resonator quality factor changes even in the case of a purely elastic sensor–sample interaction. This is due to the effective mechanical imbalance of the tuning fork prongs induced by the sensor–sample force gradient, which in turn has an impact on dissipation through the attachment of the resonator base. This effect may yield a measured dissipation signal that can be different from the one exclusively related to the dissipation between the sensor and the sample. We also find that there is a second-order term in addition to the linear relationship between the sensor–sample force gradient and the resonance frequency shift of the tuning fork that is significant even for force gradients usually present in atomic force microscopy, which are in the range of tens of N/m.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.