Abstract

Because neurosurgery and orthopedic surgery often require the drilling of bones, this paper attempts to solve the problem of bone drilling. The goal is to realize a control system that drills with a contact drilling force and can automatically stop drilling at the moment of breaking through. In particular, both the dc drilling motor control and the feed rate control are driven by force control in order to accommodate the system impedance. Moreover, breakthrough detection comes from to the threshold information of the thrust force as well as the trend of both the drilling torque and the feed rate. The proposed technique was experimentally verified through the drilling of pig bones, and the results are in accordance with the theoretical model both in the bone-drilling process and at the point of breakthrough.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.