Abstract

PurposeThe aim of this study was to evaluate the distribution of deformation stresses in customized and non-customized plates during simulated advancement genioplasty, using the finite element method. MethodsA customized plate (Traumec) was developed with 4.75 mm advancement, in ASTM F67 Grade 2 titanium, with four screws. Non-customized (standard) plates with 6 mm advancement (Stryker with six screws and Osteomed with four screws) were used for comparison. All the screws presented the same length (10 mm) and fixation system (2.0). The Traumec and Osteomed plates were fixed with two screws in the mandible, and another two in the segment, whereas the Stryker plate was fixed with three screws in the mandible, and another three in the segment. Six virtual models were generated in a computer-aided design program (Rhinoceros), in which the advancement and insertion of the plates were evaluated. All the plates were submitted to application of perpendicular and oblique forces of 5 N in the chin region. ResultsThe Osteomed plate showed the highest stress value (1506 MPa), and the Traumec, the lowest stress value (560.20 MPa). The Stryker plate showed higher stress values for the segment screws than for the mandibular screws, unlike the other plates. ConclusionsThe customized Traumec plate showed better deformation stress distribution and plate/segment stabilization when submitted to advancement genioplasty.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.