Abstract

We describe the design, construction details, and performance characteristics of an exposure system designed to provide very well controlled extremely-low-frequency magnetic field exposures of in vitro samples. This system uses Helmholtz coils placed inside temperature-controlled mu-metal chambers to provide simultaneous ac and dc field exposures at any relative angle with minimal residual background field. The system has both exposed and sham-exposed chambers and is operated under computer control in such a way as to ensure blind exposure of samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.