Abstract

This work describes a device for electrochemical enzyme-linked immunosorbent assay (ELISA) designed for low-resource settings and diagnostics at the point of care. The device is fabricated entirely in hydrophobic paper, produced by silanization of paper with decyl trichlorosilane, and comprises two zones separated by a central crease: an embossed microwell, on the surface of which the antigen or antibody immobilization and recognition events occur, and a detection zone where the electrodes are printed. The two zones are brought in contact by folding the device along this central crease; the analytical signal is recorded from the folded configuration. Two proof-of-concept applications, an electrochemical direct ELISA for the detection of rabbit IgG as a model antigen in buffer and an electrochemical sandwich ELISA for the detection of malarial histidine-rich protein from Plasmodium falciparum (Pf HRP2) in spiked human serum, show the versatility of this device. The limit of detection of the electrochemical sandwich ELISA for the quantification of Pf HRP2 in spiked human serum was 4 ng mL(-1) (10(2) pmol L(-1)), a value within the range of clinically relevant concentrations.

Highlights

  • The present manuscript describes a proof-of-concept prototype for a low-cost, easy-tofabricate, portable and sensitive device for performing electrochemical enzyme-linked immunosorbent assay (ELISA)

  • The analytical device we describe is fabricated entirely on hydrophobic paper and comprises two zones separated by a central axis: an embossed microwell, on the surface of which the antigen immobilization and recognition events occur, and a detection zone where the electrodes are printed; the two zones are brought into contact by folding the device along the central axis, at which point when the analytical signal is recorded

  • This study describes a device for electrochemical ELISA designed for low-resource settings

Read more

Summary

Published Version Citable link Terms of Use

Glavan, Ana C., Dionysios C. Christodouleas, Bobak Mosadegh, Hai Dong Yu, Barbara S. Smith, Joshua Lessing, M. Teresa Fernández-Abedul, and George M. Whitesides. 2014. “ Folding Analytical Devices for Electrochemical ELISA in Hydrophobic R H Paper .” Analytical Chemistry 86 (24) (December 16): 11999–12007. doi:10.1021/ac5020782. This article was downloaded from Harvard University’s DASH repository, and is made available under the terms and conditions applicable to Open Access Policy Articles, as set forth at http:// nrs.harvard.edu/urn-3:HUL.InstRepos:dash.current.terms-ofuse#OAP

EXPERIMENTAL SECTION
RESULTS AND DISCUSSION
CONCLUSION

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.