Abstract

This paper demonstrates a method to significantly enhance the detection efficiency of filter-based spectral sensors without the use of additional dichroic optics for spectral preselection. The fundamental principle is that light reflected from one interference filter or filter segment can be used consecutively, reducing the overall system losses. The proof-of-concept is presented using two compact optical modules. The first module uses 10 individual filters between 520 and 800nm, and the second is capable of continuous spectrum acquisition between 450 and 825nm using a linear variable filter (LVF) as a key element. An efficiency increase factor of up to approximately 100 compared to a common system, where the entire LVF is directly illuminated, was demonstrated.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.