Abstract

The focused ion beam lift-out technique for scanning electron microscope (SEM) and transmission electron microscope (TEM) sample preparation was shown to be applicable to copper/low-k dielectric semiconductor technology. High resolution SEM, TEM, and scanning transmission electron microscope analyses were performed on metal contacts and resist vias with no evidence of the interface damage or metal smearing commonly observed with mechanical polishing. Ion milling of the sample ex situ to the substrate provided decoration and adjustment of the exposed plane of the section when necessary for SEM analysis.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.