Abstract
Abstract Reduced feature dimensions in microelectronic devices has led to a growing reliance on the focused ion beam (FIB) for scanning electron microscopy (SEM) and transmission electron microscopy (TEM) specimen preparation. The introduction of copper for use in electrical interconnects will increase this reliance. Copper, much more so than aluminum, tends to smear when conventional polishing techniques are employed, rendering mechanical polishing unsuitable for quality specimen preparation. The sputtering characteristics of copper differ greatly from the aluminum based materials currently in use. Copper exhibits a grain orientation dependent sputtering rate that tends to produce uneven FIB polished surfaces. The sputtering rate for copper is also much higher than the silicon and dielectric materials that must be simultaneously removed. These differences in sputtering rates lead to the formation of curtains and ledges that also affect specimen quality. Another FIB induced artifact was observed when preparing SEM specimens (Figure 3).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.