Abstract
This paper offers a technique for non-contact assessment of material volume loss at the femoral head taper interface of modular replacement hips with novel use of the focus variation principle. A novel 3D to areal data conversion technique also allows powerful areal data analysis techniques to be applied to point cloud (3D) measurements for volume loss optimisation. Accurate characterisation of retrieved femoral head tapers is important to assess their in vivo performance particularly with respect to the bio toxicity of wear and corrosion products.A cohort of 8 retrieved femoral heads showing a wide range of degradation (0.5mm3 < volume loss < 12mm3) [1] was selected for the development of this technique. Using this common cohort of retrieved hips, the current technique was benchmarked against the well proven roundness measurement machine (RMM) method. This existing technique generates areal (2.5D) data and exploits a range of existing areal analysis techniques to optimise volume loss assessment. For benchmarking continuity and to exploit the same areal techniques, volume loss analysis for the current technique was carried out using the software written for the existing RMM method.The focus variation instrument’s integrated language was used to write script to convert (un-wrap) the taper surface 3D data into areal format.The current method shows a mean absolute difference in volume loss of 14% (-12% signed) from that of the benchmark with a range of 1% to 27%. The spread of measured values is significantly higher for the current method than for the benchmark. However, it is noted that replication can offer the advantage of capturing the whole taper surface on some taper types where physical access is limited for a stylus based roundness method.The current technique is also compared to the existing Redlux™ technique in which replicated female tapers are measured using a confocal instrument. The current technique is shown to have comparable performance to the Redlux™ technique but offers a more sophisticated methodology for volume loss analysis. In addition the current technique offers new instrumentation and analysis tools to the field.Small uncontrolled casting variations are noted in the current technique, resulting in poorer performance with small volume loss samples where the influence of this effect is most pronounced. However, given the simplified assumptions of the volume loss calculation where results may be skewed by deposits, some uncertainty will be evident with any approach.
Highlights
Learmonth, Young and Rorabeck [1] refer to the total hip replacement procedure as the operation of the century
The focus variation (FV) volume loss data is expressed with respect to the roundness measurement machine (RMM) data in both mm3 and percentage terms and ‘absolute’ mean values are quoted
Repeatability data is given for the RMM technique, thought this data is from a sample not measured in the current study and has a lower volume loss than the samples reported for the current study
Summary
Young and Rorabeck [1] refer to the total hip replacement procedure as the operation of the century. The modular taper interface design allows individual components to be selected to allow joint geometry to be tailored to individual recipients during surgery. These taper interface surfaces lack fully developed standardisation and tolerancing for; length, diameter at depth, cone angle, roundness, straightness and surface finish. Several studies have reported significant wear and the formation of corrosion products at these taper interfaces [3,4,5] see Figure 2. The current work focuses on the “large head metal on metal” (LHMoM) design which are more prone to taper corrosion and wear due to increased bending moments at the stem/head taper interface. Goldberg et al [8] detail mechanically-assisted crevice corrosion, and the similar mechanism of fretting corrosion is offered in [9]
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.