Abstract

Many mammography units now have microfocal spots ranging in size from 0.09 to 0.4 mm for magnification radiography. On site measurements of these focal spots are not possible without invasive procedures if National Equipment Manufactures Association (NEMA) specifications are to be followed. This paper describes a method to perform such measurements expeditiously (1.5 h) and noninvasively. We describe how this method differs from NEMA test specifications, and report the test results of 22 units from ten manufacturers. Six of the focal spots were larger than the manufacturer's specifications which were based on current NEMA standards. Emphasis is placed on the need for verifiable standards for mammography x-ray systems.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.