Abstract

A novel coaxial biopsy system powered by a handheld drill has recently been introduced for percutaneous bone biopsy. This technical note describes our initial experience performing fluoroscopy-guided vertebral body biopsies with this system, compares the yield of drill-assisted biopsy specimens with those obtained using a manual technique, and assesses the histologic adequacy of specimens obtained with drill assistance. Medical records of all single-level, fluoroscopy-guided vertebral body biopsies were reviewed. Procedural complications were documented according to the Society of Interventional Radiology classification. The total length of bone core obtained from drill-assisted biopsies was compared with that of matched manual biopsies. Pathology reports were reviewed to determine the histologic adequacy of specimens obtained with drill assistance. Twenty eight drill-assisted percutaneous vertebral body biopsies met study inclusion criteria. No acute complications were reported. Of the 86% (24/28) of patients with clinical follow-up, no delayed complications were reported (median follow-up, 28weeks; range 5-115weeks). The median total length of bone core obtained from drill-assisted biopsies was 28mm (range 8-120mm). This was longer than that obtained from manual biopsies (median, 20mm; range 5-45mm; P=0.03). Crush artifact was present in 11% (3/28) of drill-assisted biopsy specimens, which in one case (3.6%; 1/28) precluded definitive diagnosis. A drill-assisted, coaxial biopsy system can be used to safely obtain vertebral body core specimens under fluoroscopic guidance. The higher bone core yield obtained with drill assistance may be offset by the presence of crush artifact.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call