Abstract

PurposeTo quantify and categorize fluoroscopically-guided procedures with radiation doses exceeding 5000 mGy reference point air kerma (Ka,r). Ka,r > 5000 mGy has been defined as a “significant radiation dose” by the Society of Interventional Radiology. Identification and analysis of interventions with high radiation doses has the potential to reduce radiation-induced injuries.Materials and methodsRadiation dose data from a dose monitoring system for 19 interventional suites and 89,549 consecutive patient encounters from January 1, 2013 to August 1, 2019 at a single academic institution were reviewed. All patient encounters with Ka,r > 5000 mGy were included. All other encounters were excluded (n = 89,289). Patient demographics, medical specialty, intervention type, fluoroscopy time (minutes), dose area product (mGy·cm2), and Ka,r (mGy) were evaluated.ResultsThere were 260 (0.3%) fluoroscopically-guided procedures with Ka,r > 5000 mGy. Of the 260 procedures which exceeded 5000 mGy, neurosurgery performed 81 (30.5%) procedures, followed by interventional radiology (n = 75; 28.2%), neurointerventional radiology (n = 55; 20.7%), and vascular surgery (n = 49; 18.4%). The procedures associated with the highest Ka,r were venous stent reconstruction performed by interventional radiology, arteriovenous malformation embolization performed by neurointerventional radiology, spinal hardware fixation by neurosurgery, and arterial interventions performed by vascular surgery. Neurointerventional radiology had the highest mean Ka,r (7,799 mGy), followed by neurosurgery (7452 mGy), vascular surgery (6849 mGy), and interventional radiology (6109 mGy). The mean Ka,r for interventional radiology performed procedures exceeding 5000 mGy was significantly lower than that for neurointerventional radiology, neurosurgery, and vascular surgery.ConclusionsFluoroscopically-guided procedures with radiation dose exceeding 5000 mGy reference point air kerma are uncommon. The results of this study demonstrate that a large proportion of cases exceeding 5000 mGy were performed by non-radiologists, who likely do not receive the same training in radiation physics, radiation biology, and dose reduction techniques as radiologists.

Highlights

  • Invasive, image-guided interventional procedures have expanded the scope of medical practice across numerous domains of medicine, owing to the demonstrated benefit to patients (Heilmaier et al 2015)

  • The procedures associated with the highest Ka,r were venous stent reconstruction performed by interventional radiology, arteriovenous malformation embolization performed by neurointerventional radiology, spinal hardware fixation by neurosurgery, and arterial interventions performed by vascular surgery

  • The results of this study demonstrate that a large proportion of cases exceeding 5000 mGy were performed by non-radiologists, who likely do not receive the same training in radiation physics, radiation biology, and dose reduction techniques as radiologists

Read more

Summary

Introduction

Image-guided interventional procedures have expanded the scope of medical practice across numerous domains of medicine, owing to the demonstrated benefit to patients (Heilmaier et al 2015). In 2009, the Society of Interventional Radiology (SIR) set forth recommendations regarding radiation dose management related to interventional radiologic procedures (Stecker et al 2009). These guidelines defined a significant radiation dose threshold as any of the following: PSD > 3000 mGy, Ka,r > 5000 mGy, DAP > 500 Gy·cm, or FT > 60 min (Stecker et al 2009). These thresholds were established to identify patients who require clinical follow-up for potential deterministic radiation-induced injury

Objectives
Methods
Results
Discussion
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.