Abstract

Percutaneous fixation of pelvic and acetabular fractures are technically demanding procedures, and high rates of screw misplacement and potential neurovascular complications have been reported.One hundred and sixty two screws from a prospectively collected database were analysed to evaluate the accuracy of a fluoroscopically assisted computer navigated technique to insert a cannulated screw to treat pelvic and acetabular fractures. Actual screw position and trajectory with the intraoperative surgical plan stored in the navigation computer.The actual screw position differed from the surgical plan by a mean of 3.9mm, with a mean 1.4 degree difference in screw trajectory. Post operative CT analysis of patients showed 10 screws perforated cortical bone.Our results show that the use of computer navigation can aid in the accurate placement of percutaneous screws along a predefined plan. It is still possible to incorrectly place a screw and great care needs to be taken with the surgical plan and also to understand the complex anatomy of the bony pelvis.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.