Abstract

The fluorometric scopoletin-horseradish peroxidase method has been modified for field determinations of hydrogen peroxide concentrations in groundwaters. Standard additions calibration compensates for background fluorescence and inconsistent stoichiometry of the fluorescence quenching reaction due to interferences by the matrix. The detection limit, defined as the blank plus three standard deviations, ranged from 3.6 to 44.6 nM. However, this limit was more an indication of the difficulty of preparing peroxide-free water than the actual limit imposed by the sensitivity of the method for the peroxide contamination introduced with the reagents. For 111 field determinations the weighted average (uncorrected) hydrogen peroxide concentration was 20.2 nM and the pooled standard deviation was 7.7 nM. The average of 45 field blanks was 7.8 nM with a pooled standard deviation of 5.2 nM. At nanomolar concentration levels, it is essential that samples are analyzed for H/sub 2/O/sub 2/ in the field. Storage periods exceeding 1 h caused serious errors and irreproducible results.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.