Abstract
We investigated the effects of fast recovery (FR) to increase the sensitivity of fluorine-19 ((19)F) fast spin echo (FSE) in mapping 5-fluorouracil (5-FU) and its metabolites. We added an additional 90 degrees pulse (which flips back longitudinal magnetization at the end of the sequence) to the chemical shift selective (19)F FSE pulse sequence. In 5-FU solution, FR remarkably improved the signal-to-noise (S/N) ratio of (19)F 5-FU images, having higher effects with shorter repetition time and smaller echo train numbers. In animal studies, FR produced a conspicuous increase in (19)F signals in the urinary bladder. FR effects for (19)F signals in the liver were smaller than those in other organs but still substantial. Utilization of FR in (19)F FSE images promises more sensitive observation of (19)F metabolite maps of 5-FU and other (19)F-containing compounds that have relatively long relaxation times.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.