Abstract
A fluorimetric procedure for the determination of aluminium with matrix removal in drinking water is proposed. The system is based both on the solid phase extraction of aluminium on a new chelating resin (XAD-4 modified by grafting salicylic acid) and the fluorimetric detection of a complex formed between 8-hydroxyquinoline-5-sulfonic acid (HQS) and Al(III), after elution of the resin by hydrochloric acid. The sorption and elution of aluminium were studied in both competitive and noncompetitive conditions, varying pH, flow-rates, volume and concentration of reagents, as well as time contact. The optimised procedure allows determination of Al3+ at the sub-ppb level (LOD: 0.2 microg L(-1) for 1 ml of sample) within a working range of 0.2-500 microg L(-1). The analytical procedure was successfully employed for the determination of aluminium in drinking water during and after flocculation/coagulation treatment processes.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.