Abstract

The accuracy of the fluorescent (FM) and radioactive microsphere (RM) techniques is similar in acute experiments but has not been established in chronic experiments. In the present study various combinations (at least pairs) of FM and/or RM labels were injected simultaneously between 2 mo and 5 min before each animal was killed. Blood flow was determined in many organs. Intramethod mean difference and variation did not change over time for FM but increased significantly for RM (from 1.8 +/- 1.4 to 25.6 +/- 21.8% and from 4.4 +/- 3.2 to 32.4 +/- 23.0% at 5 min and 2 mo, respectively). Also the FM-RM intermethod mean difference and variation increased (from -0.5 +/- 8.5 to 40.8 +/- 23.8% and from 23. 6 +/- 4.6 to 71.8 +/- 34.3%, respectively). After 2 mo, blood flow estimations were 20-50% lower with the various RM, whereas brain and liver blood flow values varied even more between isotopes. Underestimation started within 1 day for 51Cr and within 2 wk for 141Ce, 95Nb, and 85Sr. We conclude that FM are superior to RM for blood flow determination in experiments lasting longer than 1 day, presumably because of leaching of isotopes from RM.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.