Abstract

In this article, Fluorescence spectroscopy has been employed for the assessment of microbial load and it has been compared with the gold standard colony forming unit (CFU) and optical density (OD) methods. In order to develop a correlation between three characterization techniques, water samples of different microbial loads have been prepared by UVC disinfection method through an indigenously developed NUVWater sterilizer, which operates in close cycle flow configuration. A UV dose of 58.9mJ/cm2 has been determined for 99.99% disinfection for a flow rate of 0.8l/min. The water samples were excited at 270nm which results in the tryptophan like fluorescence at 360nm that decreases gradually with increase of UVC dose, indicating the bacterial degradation and it has been confirmed by OD and CFU methods. In addition, it has been proved that a close cycle water flow around UV lamp is imperative so that an appropriate dose must be delivered to microorganisms for an efficient disinfection. It has been found that due to the sensitive nature of Fluorescence spectroscopy, it yields immediate results, whereas, for CFU and OD methods, water samples needs to be inoculated for 24h. Fluorescence spectroscopy, therefore, provide a fast, online, reliable and sensitive method for the monitoring of pathogenic quantification in drinking water.