Abstract

This paper describes a fluorescence technique for the detection of sub-ppm-concentrations of formaldehyde in air at atmospheric pressure. The aldehyde is excited by a frequency doubled tunable dye laser covering the 320-345-nm wavelength range. The fluorescence intensity, measured at right angles through a 400-nm cutoff filter, is proportional to the formaldehyde concentration in air. Interferences by NO(2) and SO(2) are negligible. The detection limit of the system is 5 pphm at present, in good agreement with an estimate of the fluorescence yield in 1 atm of air. A considerable increase of the sensitivity is possible, and the use of this technique for measuring formaldehyde in ambient air in the ppb concentration range is projected.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.