Abstract

The fluorescence of tifluoperazine at EX:300 nm and EM:485 nm was changed into non-fluorescent substance in the presence of bromate under acidic conditions, and the fluorescence detection-FIA of trace amounts of the bromate was developed using this phenomenon. In the concentration range of 1-15 μg L -1 , the bromate analysis was able to obtain a good straight line and eighteen samples per hour were measured. For 7.5 μg L -1 of bromate, the relative standard deviation(RSD) was 2.03 %(n=5) and the detection limit (3σ) was 1.47 μg L -1 .

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.