Abstract

The diameter of an X-ray beam was determined, using the knife-edge technique, widely applied for beam profiling, by taking advantage of the fluorescence emission generated by the X-ray beam. The knife-edge has to be appropriate to the configuration of the device, in our case a double-material target made of plastic and cardboard was scanned in a transversal plane compared to the beam propagation direction. Along the scanning axis, for each position, the intensity of the Kα line of chlorine was recorded. The first derivative of the intensity evolution as a function of the edge position, fitted by a Gaussian function, makes it possible to obtain the beam diameter along the scan direction. We measured a slightly elliptic diameter close to 3mm. In this note we underline the significance of the knife-edge technique which represents a useful tool, easy to be set up, to control X-ray beam dimensions in portable devices often routinely used by non-specialists.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.