Abstract
FLUKA-based Monte Carlo calculations of microdosimetric distributions in water phantom involving a walled spherical Tissue-Equivalent Proportional Counter filled with tissue-equivalent propane gas have been studied for an indigenously developed telecobalt machine. The simulated site size considered in the study was 2 μm. In the Monte Carlo calculations, field size was varied from 10 cm × 10 cm to 35 cm × 35 cm and the depth was varied as 5-20 cm. The study also includes calculation of microdosimetric distributions with a 30° wedge filter. The efficiency of the calculations was improved up to a factor of 26 by choosing appropriate cut off values for production and transport of electrons. The calculated microdosimetric distributions of telecobalt machine is distinctly different from that of a bare 60Co source which is attributed to the influence of scattered photons from the machine head and the water phantom.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.