Abstract
Coriolis mass flowmeter (CMF) is widely used in the industrial field. In mass flow measurement, there are many impurities in measured fluids that will adhere to the inner wall of the vibrating tube of CMF. The vibration characteristics of CMF would change due to the structural change, i.e., wall clung state, which will generate the wall clung state fault. In this paper, aiming at the wall clung state fault of CMF, the finite element model of CMF is established based on ANSYS. The velocity distribution of fluid in the vibrating tube of CMF is analyzed, considering the fluid–structure interaction. The location of the wall clung state in a vibrating tube is determined. Then, the fault model is established. The mechanism of the vibration transmission characteristics outwards of CMF caused by the wall clung state is analyzed by harmonic response analysis. Finally, the failure mode of CMF is investigated.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.