Abstract
An investigation was undertaken into the relationship between the wet length of a standard Schirmer paper and the fluid weight in the 5 mm tip after a 5-minute moistening period. Immediate weighing after moistening showed that it is unreliable to infer a particular fluid weight in the 5 mm tip from a known wet length, and it is concluded that such papers will lead to error in lysozyme tests. If the papers are inserted for a period within sealed envelopes, there is found to be a close relationship between wet length and fluid weight in the tip, the papers reaching a stable state at 5 hours. Soaked Schirmer papers were shown to given particularly inconsistent results.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.