Abstract
The effect of hemodynamic load on various stent-graft designs used for endovascular aneurysm repair (EVAR) in cardiovascular treatments is studied using a numerical fluid-structure interaction (FSI) analysis that couples computational fluid dynamics (CFD) and finite element analysis (FEA). Radial displacements, mechanical stresses, wall shear stress and wall compliance quantities are evaluated for four stent materials and one graft material. The strut thickness is varied from 0.3 mm to 1 mm. The materials are assumed linearly elastic and isotropic while blood is assumed as a Newtonian and incompressible medium with a pulsatile and turbulent flow profile. Time dependent pressure conditions are assumed at the inlet and outlet. Results are benchmarked against a study taken from the literature and indicate that the stent material and the strut thickness greatly influence the mechanical behavior of the structure. This computational study will serve as an additional tool to vascular surgeons when assessing the choice of material and design for stent-graft recipients.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of the Mechanical Behavior of Biomedical Materials
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.