Abstract
To compare infusion and extrusion fluid volumes with three 25-gauge vitrectomy systems. The infusion and aspiration rates of 25-gauge systems from Alcon (Fort Worth, TX, USA), Bausch & Lomb (St Louis, MO, USA), and the Dutch Ophthalmic Research Centre (DORC, Zuidland, the Netherlands) were measured in vitro using balanced saline solution with different heights (40, 60, 80, 100, 120 cm) of infusion bottle, cutter velocities (800, 1100, 1500 cuts/ min) and aspiration powers (100, 200, 300, 400, 500 mmHg). The infusion flow velocities differed among the systems (p < 0.001 for all comparisons [p-values 0.0009-0.0001]). The aspiration rates of the Alcon and Bausch & Lomb systems were lower than that of the DORC system (p < 0.02 for all comparisons [p-values 0.018-0.002]). Infusion and extrusion fluid rates differ among 25-gauge vitrectomy systems. These results may help to optimize ideal aspiration and infusion parameters among 25-gauge vitrectomy systems used to treat vitreoretinal diseases.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.