Abstract
Assessing portal vein (PV) hemodynamics is an essential part of liver disease management/liver surgery, yet the optimal methods of assessing intrahepatic PV flow have not yet been established. This study investigated the usefulness of 7-Tesla MRI with hemodynamic analysis for detecting small flow changes within narrow intrahepatic PV branches. Flow data in the main PV was obtained by two methods, two-dimensional cine phase contrast-MRI (2D cine PC-MRI) and three-dimensional non-cine phase contrast-MRI (3D PC-MRI). Hemodynamic parameters, such as flow volume rate, flow velocity, and wall shear stress in intrahepatic PV branches were calculated before and after a meal challenge using 3D PC-MRI and hemodynamic analysis. The hemodynamic parameters obtained using 3D PC-MRI and 2D cine PC-MRI were similar. All intrahepatic PV branches were clearly depicted in eight planes, and significant changes in flow volume rate were seen in three planes. Average and maximum velocities, cross-sectional area, and wall shear stress were similar between before and after a meal challenge in all planes. 7-Tesla 3D PC-MRI combined with hemodynamic analysis is a promising tool for assessing intrahepatic PV flow and enables future studies in small animals to investigate PV hemodynamics associated with liver disease/postoperative liver recovery.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: HPB : the official journal of the International Hepato Pancreato Biliary Association
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.