Abstract

The aim of this study was to evaluate whether confocal laser scanning tomographic measurements of the optic nerve head depend on the time of day when the examinations are performed. Thirty-two eyes of 20 subjects repeatedly underwent optic disc confocal laser scanning tomography using the Heidelberg Retina Tomograph I. Operated by a single examiner, measurements were performed at 10 a.m. and 4 p.m. on the same day. Comparing both measurements, differences in optic cup area (0.119+/-0.126 mm(2); p=0.89), neuroretinal rim area (0.120+/-0.131 mm(2); p=0.86), cup volume (0.069+/-0.086 mm(3); p=0.86), rim volume (0.083+/-0.090 mm(3); p=0.51), and retinal nerve fiber layer thickness (0.075+/-0.063 mm; p=0.31) were statistically not significant. Intraocular pressure differences between morning and afternoon ranged between 0 and 14 mmHg. Confocal laser scanning tomographic measurements of the optic nerve head are not markedly influenced by the time of day.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.