Abstract

A study was conducted to measure in vitro the procoagulant properties of platelets induced by flow through Carbomedics bileaflet and Bjork-Shiley monoleaflet mechanical heart valves (MHVs). Valves were mounted in a left ventricular assist device, and platelets were circulated through them under pulsatile flow. Platelet activation states (PAS) were measured during circulation using a modified prothrombinase method. Computational fluid dynamics (CFD) simulations of turbulent, transient, and non-Newtonian blood flow patterns generated by the two valve designs were done using the Wilcox k - w turbulence model, and platelet shear-stress histories (the integral of shear-stress exposure with respect to time) through the two MHVs were calculated. PAS measurements indicated that the bileaflet MHV activated platelets at a rate more than twice that observed with the monoleaflet MHV. Turbulent flow patterns were evident in CFD simulations for both valves, and corroborated the PAS observations, showing that, for particles close to the leaflet(s), shear-stress exposure in the bileaflet MHV can be more than four times that in the monoleaflet valve.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.