Abstract
An improved apparatus for measuring thermal diffusion in laminar flow is described. The flow cell is calibrated with 0.05 m aqueous cadmium sulphate chosen for its known thermal diffusion properties and its suitability for conductimetric analysis. Agreement with theoretical predictions (the phenomenological equations are solved in part 2 of this paper) is found to be excellent over the entire approach to the diffusional steady state. Convection effects, which are the main source of error in the conventional pure Soret effect cell, appear to be negligible. This fact makes it possible to operate the flow cell close to the steady state where the relation between the Soret coefficient and the measured separation is independent of the calibration. Versatility is a further inherent advantage in the flow cell. Large samples of liquid are collected which can be analyzed at leisure after termination of the experiment by any analytical technique appropriate to the system under investigation. It is therefore possible to examine systems of widely different physical properties without material alteration to the cell.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.