Abstract
Flow visualization analysis was applied to the Baylor/Miwatec centrifugal artificial heart to evaluate its fluid dynamic characteristics regarding antithrombogenicity. An eccentric vortex was found both in the upper and the lower gaps of the impeller, which is supposed to be caused by the eccentric inlet port. Therefore, one-way flow toward the outlet is formed and washes the pivot. The combination of an eccentric vortex and a pivot bearing that is washed is unique to the Baylor/Miwatec pump. For the male pivots exposed to periodic wash, the minimum shear rate around the bottom pivot was estimated to be 650/s, which is higher than the threshold for thrombus formation shown by other studies. The wall shear rate at the impeller bottom surface was found to be larger in the top contact mode than in the bottom contact mode.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.