Abstract

A method for flow-through liquid scintillation counting in reversed-phase liquid chromatography (RP-LC) based on the continuous extraction of aqueous column eluates with a water-immiscible liquid scintillator is evaluated in terms of sensitivity, reproducibility and extra-column band broadening. 3H- and 14C-labelled phenylthiohydantoinamino acid derivatives of widely different polarity serve as model compounds. For extractable derivatives, counting efficiencies of over 30 and 80% can be obtained for 3H and 14C, respectively. The reproducibility and extra-column band broadening depend on the mixing ratio of scintillator and LC eluent; relative standard deviations (peak areas) of less than 3% can be obtained. The sensitivity of flow-through counting can be increased at least 150-fold by storing the segmented scintillator/eluate stream in a capillary storage loop. After the separation is complete, the stream is re-introduced into the radioactivity detector at reduced flow-rates to increase the mean residence time, i.e., the counting time, in the detector.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.