Abstract

We report experimental results from a flow-through integrating cavity absorption meter. The operating range of the device is from 0.004 m(-1) to over 80 m(-1) of absorption. Absorption coefficients have been measured with 8% or less change in the presence of over 200 m(-1) of scattering in the medium. The instrument signal has been shown to be independent of flow rate up to 20 liters/min and thus independent of turbulence. This large operational range along with the ability to measure absorption independently of adverse scattering affects allows the instrument to be utilized in a wide range of environmental conditions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.