Abstract

Infusion systems are widely used in clinical medicine. Intravenous infusion therapy must be monitored to ensure patient safety. We proposed a compact flow rate sensor device based on the time-of-flight method. This device included one ceramic heater and two infrared sensors. Practical sensor prototypes were fabricated and characterized. The response time was 30 s. The sensor range was estimated to be 33 dB from μL/min to tens of mL/min, covering almost the entire usage range This flow rate sensor can be applied to common infusion tubes. Through the use of a mobile phone app, detailed information can be presented in real time.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.