Abstract

A flow analysis (FA) system with spectrophotometric detection for Al determination in hemodialysis solutions was developed. The method was based on the reaction of Al with eriochrome cyanine R (ECR). The complex formed associated with cetyltrimethylammonium bromide (CTAB) — a cationic surfactant, which showed enough sensitivity to execute the direct analyte determination. All interferences were eliminated with the matrix matching calibration. The system presented the following analytic parameters: sensitivity ( m) of 8.10 × 10 −4 L μg −1, limit of detection (LOD) of 3.24 μg L −1 (3 σ), linear correlation coefficient of 0.9966 and linear range response from 10.8 to 650 μg L −1. The accuracy of the proposed method was checked by comparison with electrothermal atomic absorption spectrometry (ET-AAS) method. There were no differences among the results obtained from both methods, at a confidence level of 95% (paired t-test). Recovery tests were also made, values obtained were from 90.4 to 109 of recovery for Al-spiked samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.